Storing a Vscan Air probe requires careful consideration of several factors to maximize the lifespan of the probe battery.
Longer term storage - more than one month
If there is a need to store a Vscan Air probe for a longer period, it is recommended to utilize ‘Storage mode’ available via the Vscan Air app Menu.
‘Storage Mode’ allows the user to keep a probe stored for up to 36 months without significant probe battery degradation. If the probe must be available on short notice, it can easily be removed from ‘Storage mode’ by placing it onto a Vscan Air charger.
Short term storage
If selecting to store the probe for a short period without utilizing Storage mode it is recommended to:
• Keep the Vscan Air probe OFF when storing it or select to use ‘Storage mode’ even if storing the probe for less than one month (see Putting the probe into Storage mode).
• Store the Vscan Air probe in a cool, dry place. Aim for temperatures between 0°C to 20°C and remember that colder is better, for example: 10°C is better than 20°C.
• Although a Vscan Air probe carrying a fully charged battery can be stored up to 3 months without needing to be recharged, it is recommended to charge the probe on a monthly basis to make sure that it carries enough power for a scan session if there is a sudden need for starting using the probe.
• If a dedicated display device is stored with the probe keep it OFF when storing.
• Re-charge the display device at the intervals required to make sure that the device will be able to keep the Vscan Air app actively running for a minimum of 2 hours. Follow the display device manufacturer recommendations regarding the display device battery, battery charging and battery health
Important guidelines to follow for Storage
• Store probe in a cool, dry place. Aim for temperatures between 0°C to 20°C and remember that colder is better, for example: 10°C is better than 20°C.
• Make sure to re-enable probes in ‘Storage mode’ at least every 36 months to check battery capacity, charge if needed, and put probe back into ‘Storage mode’ if required.
• If storing multiple probes, it is recommended to practice a first in first out rotation when selecting a probe for use.