Probe LEDs steady blue, app never proceeds to scanning

Info
This state reflects that the network has been established between the app device and the probe. If the searching for probe screen has transitioned to the preset menu, there was an issue selecting the saved default preset. 


Solution
Select the desired preset. If the problem persists, it is an indication that the socket communication link over the network connection was lost. Reboot the probe and try again. 





Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!