Environmental Requirements

Environmental Requirements for using Vscan Air probe - Operational, Non operational, Storage and transport conditions for Vscan Air. 


Environmental Requirements for Vscan Air probe

Description

Operational

Non operational

Storage and transport

Temperature

0°C to + 35°C

- 40°C to + 70°C

- 40°C to + 70°C

Humidity

10–90%

10–95%

10–95%

Air pressure

54 kPa to 106 kPa

50 kPa to 106 kPa

50 kPa to 106 kPa


Transient Operating Conditions


Permissible transient environmental operating conditions:

  • Temperatures ranging from -20°C to 0°C or from +35°C to +50°C
  • Device will function for a minimum of 20 minutes when placed in an environment with temperatures ranging from -20°C to 0°C or from +35°C to +50°C after storage at room temperature (20 +/- 2°C)
NOTE: Limitation in the 50°C ambient environment is 'probe stops scanning' once the probe front face (applied part) temperature exceeds 43°C. To avoid this happening the following test sequence was selected and repeated 3 times (21 minutes).
1. Probe connected to Vscan Air app - ready to scan for 5 minutes
2. Scanning for 2 minutes
This test would be one example of a simulated realistic scenario at 50°C.
  • Following storage at temperatures ranging from -20°C to 0°C or from +35°C to +50°C, the device will within 10 minutes after being returned to room temperature (20 +/- 2°C), function for a minimum of 20 minutes.
NOTE: Avoid exposing the unit to saline moisture. In case of exposure to saline moisture, clean the unit as described in ‘Cleaning and Disinfection'
Image on the display device hosting the Vscan Air app is dependent on ambient light; avoid direct sunlight or reflections from other light sources on the display when scanning and reviewing images. The display viewing angle should be as small as possible.

If you are having difficulty seeing the image due to the lighting conditions try to change brightness of the display device or change your position/location of use.


Acclimatization Time


Allow the Vscan Air probe to acclimate for approximately 10 minutes prior to use, if stored at temperatures ranging from -20°C to 0°C or in temperatures ranging from +35°C to +50°C.


Allow the Vscan Air probe to acclimate for approximately 30 minutes prior to use, if stored at temperatures ranging from -40°C to -20°C or in temperatures ranging from +50°C to +70°C.


Other environment: Aircraft Ambulance/ Road Ambulance


The Vscan Air ultrasound system can be used in an emergency medical services environment, including road** and air ambulances*.

  • The Vscan Air probes are not certified for being mounted or fixed inside a road- or air ambulance.
  • The Vscan Air probes can be used in emergency medical services environments.
  • The Vscan Air probes are not certified for being charged inside a road- or air ambulance.
  • The AC Adapter must not be operated above 5000m.


* The Vscan Air ultrasound system is compliant to IEC 60601-1-12 and EN13718. Additional regulations might apply

** The Vscan Air ultrasound system is compliant to IEC 60601-1-12 and EN1789. Additional regulations might apply


Disclaimer: The information herein is not intended to replace the product’s User Manual. Please consult the User Manual for comprehensive information and cautionary guidance about the product’s operation and use.

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article