Important info - Account Creation (Mar 26th): Account creation via Vscan Air application is expected to become available on Friday, Mar 27th. An update will be shared once account creation is active.
  1. Vscan Air ™
  2. Solution home
  3. Vscan Air - Maintenance & Troubleshooting
  4. Troubleshoot: Pairing & Connectivity
Open navigation

Probe LEDs turn purple(magenta), shuts down during connection

Info
The probe attempts to send a message to the app device as to why it is shutting down. But if the battery level is low as it is booting, it may end up shutting down before the connection completes, or before the message has been sent.


Solution
Place the probe on the charger and try again when sufficiently charged. 
 
If the problem persists, the probe has an issue that requires service.





Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article