Probe Overheats
WARNING: Probe is overheating and the system will shut down. If the problem persists, contact GE HealthCare.Battery Low
WARNING :Battery critically low. Connect charger immediately.Probe Voltage Critical
WARNING: Probe voltage is at critical level and will shut down. If the problem persists, contact GE HealthCareConnectivity Error
WARNING: Unable to connect. Check the internet connection and try againHandling A Defective Vscan Air Probe Or Battery
If the Vscan Air Probe is damaged and needs a replacement, make sure you clean and disinfect the Vscan Air Probe, pack it in the packing box, before shipping or returning to GE HealthCare
If the battery does not charge, or if it is damaged. Contact GE HealthCare for battery replacement.
Automatic Deactivation Of The Vscan Air App
Deactivation Soft Limit
The Vscan Air app will on a daily basis try to contact the Vscan Air Product Registration server. If the Vscan Air app is inactive and/or not connected to internet for more than 70 days you will see the following message when you try to use the app.

Reactivate Warning Message
Deactivation Hard Limit
The Vscan Air app will on a daily basis try to contact the Vscan Air Product Registration server. If the Vscan Air app is inactive and/or not connected to internet for more than 110 days the Vscan Air app is deactivated and you will see the following message when you try to use the app.

Auto Deactivated Message
NOTE: To avoid deactivation of the Vscan Air app, make sure you use the app while being connected to internet before reaching the deactivation hard limit.Disclaimer: The information herein is not intended to replace the product’s User Manual. Please consult the User Manual for comprehensive information and cautionary guidance about the product’s operation and use.