This feature allows data to be exported to a cloud based/ remote server using the DICOMweb (STOW-Store over the web) protocol.
Follow the procedure below to configure a new DICOMweb server.
1. Press Menu → Connectivity → Servers.
Server settings
2. Select Add server on the Servers screen.
Add server
3. On the “Add server” screen, select “DICOMweb server” from the drop-down list of “Server type” and enter the configuration information in the respective fields to add the DICOMweb server
Add DICOMweb server
4. “Authorization” is an optional feature for servers that need basic authentication using credentials.
5. For DICOMweb server, the only option is secure communication. So, the check box is by default enabled. See Secure DICOM for more information.
6. Press Verify server to verify communication with the DICOMweb server. “Verify server succeed” pop-up message displays if the communication with the Server is established successfully.
Press OK and then Press Add to add the DICOMweb server. If communication fails, check the server settings, and make necessary corrections.

DICOMweb server details
Once the DICOMweb server is added, you will find the server name under Servers.