Troubleshoot the Vscan Air probe

 

ProblemPossible causeSolution
1. Vscan Air probe - can not Power ONBattery is discharged and the Vscan Air probe charge level is <8%.Probe will need to charge above 8% to be powered on.
2. Vscan Air probe is not charging.Defective battery or probe hardware issueContact GE HealthCare Service 
Defective AC adapter.Contact GE HealthCare Service
Defective wireless charger.Contact GE HealthCare Service
Defective USB cable.Contact GE HealthCare Service
Mains power is down.Ensure that the charger is connected to a powered wall outlet.
Temperature is outside the specified limits.
Ensure the ambient temperature is within the specified limits. Next step
Incompatible AC adapter or cable connected to the Vscan Air wireless charger indicated by the charger LED flashing green.Connect a compatible AC adapter and/or cable to the Vscan Air wireless charger.
A USB-C to USB-C cable is usedUse a USB-A to USB-C cable
3. Parts of the image is missing when scanning.Channels are missing.Run the transducer element test.
Contact GE HealthCare Service 
4. The Vscan Air wireless charger LED is flashing or 'breathing' blue using 2 seconds per cycle.A metal obstruction is detected.
Remove the metal obstruction from the Vscan Air wireless charger.
5. Low volume high pitch sound from
probe while scanning.
 
 
 
 
 
 
Vibrations in probe electronics might
in certain setting create a low volume
high pitched noise. This sound is more
likely to appear when activating the
curved transducer than with the other
transducers. It should be considered
being natural and not an error with the
actual probe.
A subset of users may be able to
sense the sound/noise. It may cause
discomfort, headache and/or hearing
sensations. To reduce this, consider
modifying settings if possible, moving
the probe further away from the
user, and consider also using hearing
protection.

Disclaimer: The information herein is not intended to replace the product’s User Manual. Please consult the User Manual for comprehensive information and cautionary guidance about the product’s operation and use.

 


Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article