Important info - Account Creation (Mar 26th): Account creation via Vscan Air application is expected to become available on Friday, Mar 27th. An update will be shared once account creation is active.
  1. Vscan Air ™
  2. Solution home
  3. Vscan Air - Maintenance & Troubleshooting
  4. Troubleshoot: Pairing & Connectivity
Open navigation

Mobile device connects to the wrong probe

Info
When a probe has been paired with multiple app devices, the timing of when an app was launched on the device and when a probe begins searching (waiting to pair) mode can result in the selection of the wrong device when multiple users are attempting to work in proximity.



Solution
1. Close the app fully (do not just put it in the background)
2. Power off the probe
3. Move a minimum of 1.5 meters or more away from the other probes and display devices
4. Launch the app
5. After the app display shows "Searching for probe", power on the probe and hold it near the mobile device (around 10 centimeters) to ensure it will detect the correct app device as being the closest.





Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article