Important info - Account Creation (Mar 28th): Account registrations can now be made via Vscan Air Application.
  1. Vscan Air ™
  2. Solution home
  3. Vscan Air - Know Your Device
  4. Vscan Air Battery
Open navigation

Vscan Air Probe LED Indication

Vscan Air CL / SL LED Indication when the probe is turned on, off or placed on charging.


Turning the Probe ON/OFF

To turn ON the Vscan Air Probe, press and hold the power button for 1 second.

To turn OFF the Vscan Air Probe, press and hold the power button for 3 seconds.


LED states when Vscan Air CL is ON

1. Powering up

2. Booting

3. Searching

4. Display found


5. Connected

6. Active array

7. Powering OFF/Occupied

8. Error


    • A blinking white light switching between Vscan Air Probe ends indicates that the Vscan Air CL is ON and booting.
    • A blinking white light on both LEDs indicates the Vscan Air Probe is ON and searching for Vscan Air App.
    • A steady white light on both LEDs indicates that the Vscan Air Probe is ON and a display device with the Vscan Air App is found.
    • A steady blue light on both LEDs indicates that the Vscan Air Probe is ON and connected.
    • A steady blue LED on one Vscan Air Probe side indicates which transducer is active and selected. Vscan Air CL is powered ON.
    • A steady purple light on both LEDs indicates the Vscan Air Probe is ON, is occupied and shall not be interrupted. E.g. when running test or upgrading software.
    • A steady red on both LEDs indicates the Vscan Air Probe is ON and an error has been detected
NOTE: The steady blue LED will light up in a brighter blue color when scanning compared to when scanning is halted. i.e., in Freeze. Indicating that probe now saves power while not scanning. The Vscan Air probe will switch itself off when the battery capacity drops 17% from the charge level probe had when it was left connected to the app in Freeze.

Vscan Air Probe is OFF

A short press (<1 sec) on the power button while the Vscan Air probe is off will show the battery charge status via the LED lights.

LED states when Vscan Air Probe is OFF

  1. Battery level 8-90%
  2. Fully charged


1. No light indicates that the Vscan Air probe Storage mode is enabled for one of the below reasons:

  • A new not yet used Vscan Air probe arrives with Storage mode enabled. Before starting to use the probe, it has to be activated as described in the section Vscan Air Probe activation
  • The Vscan Air probe will automatically enter Storage mode when the probe charge level is <8%. The probe will need to charge to a level above 8% to accept being powered on.
  • The probe can also have been put into Storage mode as described in the section Storing devices in between uses
NOTE: To re-activate the probe, the probe needs to be placed onto the wireless charger. When the probe has been re-activated it can be turned ON and battery status can be checked via the short press (<1 sec) as described in this section.


2. A steady yellow light on both LEDs indicates a Vscan Air Probe charge level in the range 8-90%.


3. A steady green light on both LEDs indicates a Vscan Air Probe charge level in the range 90-100%


Vscan Air Probe is Charging

The Vscan Air probe will be powered off during charging. Please refer Charging the Vscan Air Probe


LED states when Vscan Air Probe is charging

  1. Battery level <8%
  2. Battery level 8-90%
  3. Fully charged
  • A blinking light on both LEDs indicates a Vscan Air Probe charging issue.
  • A steady orange/red toned yellow on both LEDs indicates the Vscan Air Probe is charging. Charge level is less than 8%, and probe will need to charge above this level to be powered on.
  • A steady yellow light on both LEDs indicates the Vscan Air Probe is charging. Charge level in the range 8-90%.
  • A steady green light on both LEDs indicates the Vscan Air Probe is charging. Charge level 90-100%.

Disclaimer: The information herein is not intended to replace the product’s User Manual. Please consult the User Manual for comprehensive information and cautionary guidance about the product’s operation and use.

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article