1. Vscan Air ™
  2. Solution home
  3. Vscan Air - Know Your Device
  4. Vscan Air Battery
Open navigation

Battery Specification

Learn about the Battery Specifications of Vscan Air CL - Charging time, Capacity and Battery lifetime.


Items

Unit

Value

Description

Basic

Voltage

mV

3600

MAX

Current

mA

889

Avg 0.71C

Battery Specification


Battery Specifications

Item

Specification

Charging time starting at 10%, charging to 90% capacity using a new battery at room temperature 20 to 25 deg. C

75 minutes

Capacity

50 minutes continuous scan (see Note)

Lifetime

At least 500 charges


In order to get maximum charging capacity with your Vscan Air CL battery, you should initially allow the battery to be fully charged and then fully discharged at least three times. Perform normal operation during these cycles. Once the initial charging/ discharging cycles are performed, the following is applicable without reducing the lifetime of the battery:

  • It is not necessary to completely discharge the battery before re-charging.
  • It is possible to stop charging the battery before it is fully charged, but the battery will then be discharged more rapidly.
  • It is possible to charge the battery several times each day, if needed.

To minimize battery performance degradation, avoid prolonged storage of the Vscan Air CL probe outside the specified temperature range.

Batteries generally degrade by aging and number of recharging cycles, and will have reduced capacity over time.


NOTE: The 50 minutes continuous scanning time is valid for a fully charged new battery using 80% B-mode and 20% color during scanning. Scanning is taking place at normal room temperature (22°C) and the factory default for the curved array Abdominal preset is utilized.



Disclaimer: The information herein is not intended to replace the product’s User Manual. Please consult the User Manual for comprehensive information and cautionary guidance about the product’s operation and use.

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select atleast one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article