Environmental Requirements for using Vscan Air CL - Operational, Non operational, Storage and transport conditions for Vscan Air.


Environmental Requirements for Vscan Air CL

Description

Operational

Non operational

Storage and transport

Temperature

0°C to + 35°C

- 40°C to + 70°C

- 40°C to + 70°C

Humidity

10–90%

10–95%

10–95%

Air pressure

54 kPa to 106 kPa

50 kPa to 106 kPa

50 kPa to 106 kPa


Transient Operating Conditions


Note: Permissible transient environmental operating conditions:
Following storage at temperatures ranging from -20°C to +50°C, the device will within 10 minutes after being returned to room temperature (20 ± 2°C), function for a minimum of 20 minutes


NOTE: Avoid exposing the unit to saline moisture. In case of exposure to saline moisture, clean the unit as described in Cleaning and Disinfection'
Image on the display device hosting the Vscan Air app is dependent on ambient light; avoid direct sunlight or reflections from other light sources on the display when scanning and reviewing images. The display viewing angle should be as small as possible.
If you are having difficulty seeing the image due to the lighting conditions try to change brightness of the display device or change your position/location of use.


Acclimatization Time

Allow the Vscan Air CL to acclimate for approximately 30 minutes, if stored at temperatures outside the transient environmental operating conditions -20°C to +50°C.


Other environment: Aircraft Ambulance/ Road Ambulance

The Vscan Air ultrasound system can be used in an emergency medical services environment, including road and air ambulances.

  • The Vscan Air CL is not certified for being mounted or fixed inside a road- or air ambulance.
  • The Vscan Air CL can be used in emergency medical services environments.
  • The Vscan Air CL probe is not certified for being charged inside a road- or air ambulance.
  • The AC Adapter must not be operated above 5000m.

Disclaimer: The information herein is not intended to replace the product’s User Manual. Please consult the User Manual for comprehensive information and cautionary guidance about the product’s operation and use.