Important info - Account Creation (Mar 28th): Account registrations can now be made via Vscan Air Application.
  1. Vscan Air ™
  2. Solution home
  3. Using Vscan Air
  4. Preparing For A Guided Procedure With Vscan Air
Open navigation

Assessing Display Device Wi-Fi Performance with Vscan Air Probe

Prior to setting up for a guided procedure, it is recommended to check if the display device being planned for use during the procedure supports a stable connection with the Vscan Air probe, by following the below steps.


Step

Step name

Description

1

Configure app settings

Set the Video duration to 1 second from Menu -> Configuration. This will help to visualize discontinuities in loop playback which indicate lost frames.

2

Connect probe and app

Connect the probe with the app

3

Select preset with color

Select Small Parts from the Shallow (Linear) presets menu. Enter color flow mode by pressing the ROI button. Adjust the color gain to 100% to visualize noise inside the color ROI.

4

Confirm/Verify

Observe the noise pattern inside the color ROI. The randomness of the pattern should not appear to pause occasionally.

5

Review cine

Press freeze. Then press the Play button to the left of the scroll bar.

6

Confirm/Verify

Observe the blue cine position indicator as it moves from left to right along the position indicator. It should move in a consistent fashion from left to right (the distance it moves with each increment appears equal).


If the display device does not pass the above criteria successfully, using an alternate display device is recommended for supporting guided procedures.




Disclaimer: The information herein is not intended to replace the product’s User Manual. Please consult the User Manual for comprehensive information and cautionary guidance about the product’s operation and use.

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article