For a few presets, there is a control available in the color flow mode for alternating between high and low velocity scales. A higher velocity scale is needed to avoid aliasing while imaging of faster (arterial) flow. A lower velocity scale optimizes the imaging of slow flow (venous).

The presets that support selection between high and low velocity scale are following:

 

Curved Array

Linear Array

Abdominal

Vascular

OB-Gyn

MSK

Vascular

Small Parts

Cardiac


To change the velocity scale, tap on the to alternate between the velocity scales. The indicates that the higher velocity scale is selected and indicating the lower scale. The velocity setting corresponding to both scales can be seen at the top of the velocity bar .

To change the velocity scale, tap on the to alternate between the velocity scales. The indicates that the higher velocity scale is selected and indicating the lower scale. The velocity setting corresponding to both scales can be seen at the top of the velocity bar .

Fast and slow flow

  1. High velocity setting
  2. Low velocity setting

Disclaimer: The information herein is not intended to replace the product’s User Manual. Please consult the User Manual for comprehensive information and cautionary guidance about the product’s operation and use.