General Imaging Adjustments

Disclaimer: The information herein is not intended to replace the product’s User Manual. Please consult the User Manual for comprehensive information and cautionary guidance about the product’s operation and use.


Freeze

To freeze an image:


1. Press the Freeze button or tap anywhere on the image area on the scan screen.

Tap or press


When in freeze mode, you can store an image or review frames in the available video buffer.


To reactivate the image:

  1. Press Freeze icon again. OR Tap anywhere on the image.


Auto freeze

If Vscan Air CL is idle for a period of time, the device enters freeze mode to minimize risk of overheating and battery drainage. Press the display to unfreeze the image and continue scanning.


Gain

Black and white gain increases or decreases the amount of echo information displayed in an image. It may have the effect of brightening or darkening the image if sufficient echo information is generated.


1. Swipe at least half a centimeter from the left or right on the scan screen to initiate a gain change.

Drag

NOTE: Small movements will be ignored to avoid unintentional activation of the gain control.


2. Move your finger to the right or left to increase or decrease gain.

Gain

The adjusted gain value is seen with the icon on the top of the screen.


Time Gain Compensation (TGC)

TGC amplifies returning signals to correct for the attenuation caused by tissues at increasing depths.


TGC sliders are spaced proportionately to the depth. The area each slider amplifies varies as well.


Turn ON or OFF the TGC Controls from the Menu screen.


The TGC controls are displayed on the scan screen as soon as you start adjusting gain by sliding your finger on any part of the image.


Adjust individual TGC sliders to adjust the intensity of echoes in a particular depth of the image.


The TGC sliders will disappear from screen after a timeout period.

TGC

  1. Landscape mode
  2. Portrait mode



NOTE: The Gain control is disabled as long as the TGC controls are active.


Depth

Depth adjusts the field of view. It increases the field of view to look at larger or deeper structures; it decreases the field of view to look at structures near the skin line.


1. Swipe your finger up/down on the screen at least half a centimeter to initiate depth

Depth adjustment

NOTE: Small movements will be ignored to avoid unintentional activation of the Depth control. 


At the start of the swipe, the depth indicator shows the depth change.


2. Change depth per cm from 5-24 cm. OR Change depth per 0.5 cm from 2-5 cm.

NOTE: When the screen is rotated from portrait to landscape, depth remains unchanged.



Zoom and Pan

Pinch on the image with two fingers to zoom.

To pan or move the zoomed area, slide your finger on the image in the direction where you want the zoomed area to move.

 

Pinch

 

An indicator will be displayed on the image whenever Zoom is activated to help recognize a zoomed image easily.

Zooming will be disabled if clicking the on screen Zoom indicator when in Zoom.

 

Zoom indicator

  1. Landscape mode
  2. Portrait mode
NOTE:  The images can be zoomed only before saving.
NOTE: While zooming, one or both edges of the image may go out of the field of view on the screen. In such cases, you will see ‘Image crop indicators’ appear on the sides where the image edge is not visualized (Refer Image crop indicator’)


Curved array transducer

  1. Zoom indicator
  2. Crop indicator


Resolution and Penetration Setting

For a few presets, there is a control available in the black and white mode for alternating between Resolution and Penetration settings.

Resolution setting is the default setting optimized for high resolution imaging balancing adequate penetration for average sized patients.

Penetration setting is useful for imaging at greater depths by optimizing visualization of structures in the far-field in larger than average or difficult to scan patients. An increase in penetration is obtained at the expense of decreased resolution.

The presets that support selection of Penetration setting are listed below.

 

Curved array

Linear array

Abdominal

Vascular

Cardiac




To switch between the ‘Resolution’ (default) and ‘Penetration’ settings, tap on the . The   icon indicates that the ‘Penetration’ setting is selected and icon indicates default or ‘Resolution’ setting 


Resolution and Penetration setting

  1. Resolution setting (default)
  2. Penetration setting

Video Buffer

Video buffer displays with an ability to scroll, play and pause the frames in the video buffer..

The indicator is placed on the last frame. 

To jump frames:

  1. Tap on the video buffer bar to view any frame.
  2. Scroll to jump multiple frames